Recall:
press the recall button, the first record of recall list turns to green, rotate the knob to select a record, and
then press the knob to perform data recall, the recalled waveform and data will be displayed in the second channel.
In CSM recall screen, rotate the knob to view CSM trend and EEG waveforms corresponding to this point, rotate
the knob also can shift the staff toward right/left, the corresponding BS, EMG and SQI value will appear above
the waveform. The EEG waveform area can display 5seconds EEG waveform from sometime.
When rotating Navigation Knob to move the mark, the moving interval is a changing value. The rule is that the
initial step is 1, after moving it towards the same direction 5 times, the interval becomes 5, and with 5 more steps
the interval becomes 10, then 20, 40 and 80. The recalled EEG waveform is shown as Figure 4.24.
Figure 4.24 CSM Recalled Waveform
Figure 4.25 ECG Recall List
Rotate the Navigation Knob to move forward or backward to review the waveform. Press the Navigation Knob to
exit the waveform recall and return to the initial waveform recall screen.
During waveform recall, the system not only displays the current recalled waveform, but also displays the lead
status, gain and filter type of the waveform and time, as shown in Figure 4.26 Waveform Recall.
Figure 4.26 Waveform Recall
Rotate the Navigation Knob to move forward or backward to review the waveform. Press the Navigation Knob to
exit the waveform recall and return to the initial waveform recall screen.