30
Figure 4.22 CSI Graphic Trend
4.6.2 Operation Instruction
Rotate the Navigation knob to select parameter and press it to view trend graph. Choose “OBSE” to view all the
parameters at some time. The value “8h”field can be selected as 8h, 24h, and 120h. Move the cursor to “Exit”
button and press it, or just press the DISP key to exit and return to initial screen.
Note: In the temperature trend graph, the Temperature 1 is dotted in white and Temperature 2 is dotted in green.
☞ Press the Print Key to print the current graphic trend.
4.7 Recall Screen
Move the gray cursor to “Recall” in the mode selection screen, and press Navigation Knob to enter into Recall
trend screen, displays in the same position shown as Figure 4.23.
Figure 4.23 Recall List
In this screen ECG waveform recall or CSI and EEG waveform recall can be selected, we will cover the each
button function in the recall list screen.
ECG button
: For selecting the type of recall data. Two options: CSM and ECG. Choose CSM button and then
press the knob to shift between CSM and ECG. If the option is “CSM”, the CSI trend and EEG waveform recall
will be displayed; If the option is “ECG”, the ECG waveform recall will be displayed and the list area will be the
ECG recall list screen (as shown in Figure 4.25). Generally the recall record is saved every one hour, if the patient
is changed less than one hour; this piece of record will be saved as a separate record.