Information
Page 3 of 5
Test Report No.KZ2022018
This report may not be published except in full unless permission for the publication of an approved extract has been obtained in writing.
Test Results
7.9.2 Penetration of filter material
Pass
1
The penetration of the filter of the particle filtering half mask shall meet the following requirements.
Sodium chloride test 95 L/min
Paraffin oil test 95 L/min
FFP1
≤20%
≤20%
FFP2
≤6%
≤6%
FFP3
≤1%
≤1%
Note 1: FFP2 respirator. Test results are shown in Annex A Table 7.9.2.
7.16 Breathing resistance
Pass
2
Classification
Maximum permitted resistance (mbar)
Inhalation
Exhalation
30 L/min
95 L/min
160 L/min
FFP1
0.6
2.1
3.0
FFP2
0.7
2.4
3.0
FFP3
1.0
3.0
3.0
Note 2:FFP2 respirator. Test results are shown in Annex A Table 7.16.
End of Test Results