! ! WARNING! !
Adjustment
If the current loop of the output signal has an impedance of 500 ohms, the power supply should
never fall below 23 V DC.
OUTPUT SIGNAL ZERO ADJUSTMENT = 4 mA
Inject nitrogen at a rate of 30 l/h.
On the X1 terminal board, place an ammeter between terminals 1 and 2 (-). With the
potentiometer Z, adjust the current to 4 mA.
With the ammeter still connected, inject the test gas at a rate of 60 l/h. Adjust the sensitivity
with the potentiometer S (Ref. 1, Fig 21). Refer to the following calibration curves for sensors
without linearization cards.
If this fails, flip the J2 jumper and begin again.