Product Manual
Technical Information
86
Factory Defaults
The N-560 is shipped with factory default settings. Authorized
technical personnel using the procedures described in the N-560
service manual can change default settings.
Table 5: Factory Default Settings
Parameter Range
Factory Default
Set
ting
%SpO2 Upper
Alarm Limit
Lower Alarm Limit plus
1 to 100%
100%
%SpO
2 Lower
Alarm Limit
20% to Upper Alarm
Limit minus 1
85%
Pulse Rate Upper
A
larm Limit
Lower Alarm Limit plus
1 to 250 bpm
170 bpm
Pulse Rate Lower
A
larm Limit
30 bpm to Upper Alarm
Limit minus 1
40 bpm
Alarm Silence Reminder On or Off On
Alarm Silence Duration Off, 30, 60, 90, 120
second
s
60 seconds
Alarm Silence
Restr
ictions
• Audible reminder
• No audible reminder
• Do not allow alarms
of
f
Audible reminder
Alarm Sound Selector 1, 2, 3 1
Alarm Volume 1 to 10 4
Data Port Baud Rate 2400, 9600, 19200 19200
Data Port Protocol 1, 2 1 (ASCII)
Display 0, 1 0 (On)
In-Sensor Trend Mode 0, 1, 2 0 (Event SpO
2)
Language English English
Pulse Beep Volume 0 to 10 4
RS-232 Nurse Call
Pr
iority
Normally high,
normally low
Normally low
SatSeconds Off, 10, 25, 50, 100 Off