The Time column represents the N-560 real-time clock. Refer to the
N-560 service manual to set the N-560 real-time clock.
Patient Data
Patient data is highlighted in the display above. Parameter values are
displayed directly beneath the heading for each parameter. In this
example, the %SpO
2 is 100, and the pulse rate is 190 beats per
minute. An asterisk indicates that the parameter is outside the set
limits. If no data for a parameter is available, three dashes (- - -) are
displayed.
PA represents pulse amplitude. The number can range from 0 to 254.
T
here are no alarm parameters for this value. It can be used for
trending information as an indication of a change in pulse volume,
relative pulse strength, or circulation.
Operating Status
The Status column indicates alarm conditions and operating status of
the N-560. In this example, Pulse High (PH) means that the pulse rate
upper alarm limit has been exceeded. The status codes are listed in
the following table. As many as four codes can be displayed at one
time in the Status column.