N-560 Operator’s Manual 59
The last data field in the top line indicates the upper and lower alarm
limits for %SpO
2 and for the pulse rate (PR), and the SatSeconds
alarm setting (OFF). The SatSeconds setting may be OFF, 10, 25, 50,
or 100 depending on the SatSeconds alarm setting. In the example
above, the low alarm limit for SpO
2 is 85% and the upper alarm limit
is 100%. Pulse Rate alarm limits are 40 to 170 bpm. SatSeconds
alarm limit is set to off.
N-560 Mode
The N-560 mode is ADULT.
Data Column Headings
Actual column headings are in the third row of the column heading
line. Patient data that is presented in the chart, from left to right, are
the time that the patient data on the row was obtained, the current
%SpO
2 value being measured, the current Pulse Rate in beats per
minute (bpm), the current Pulse Amplitude (PA), and the operating
status of the N-560.
Time