Product Manual
Printing
N-560 Operator’s Manual 55
Printing
Printing N-560 Real-Time Data
Real-time data is continuously sent to the data port on the back of
the N-560. Patient data can be obtained through the data port by
connecting to a computer or serial printer. When a real-time printout
or display is being transmitted to a printer or computer, a new line is
printed/displayed every 2 seconds. Column headings are printed/
displayed every 25 lines, or if one of the values in the column heading
is changed.
Note: If the data output stops transmitting, the N-5
60 must be
turned off and then turned back on.
When connecting the N-560
to a printer or computer, verify proper
operation before clinical use. Both the N-560 and the printer or PC
must be connected to a grounded AC outlet.
Any printer or computer connected to the N-560's data port must be
certified accor
ding to IEC Standard 60950. All combinations of
equipment must be in compliance with IEC Standard 60601-1-1
systems requirements. Anyone who connects a printer or PC to the
data output port configures a medical system and is therefore
responsible that the system complies with the requirements of system
standard IEC Standard 60601-1-1 and the electromagnetic
compatibility system standard IEC Standard 60601-1-2.
1. Data Port Connector