Product Manual

N-560 Trend
44
Trend Data
Trend data information may be retrieved or cleared through the
N-560 data port using options available in a display menu.
To access the menu options, simultaneously press the SpO2 Alarm
Limit
and Pulse Rate Alarm Limit buttons until Option 1 appears on
the display.
Then, using the Adjust Up button and Adjust Down button, you may
scroll through the available menu options as fo
llows:
#1: Trend Print
Allows printing of the existing trend data. The output may be viewed
on a PC using the Hyper Terminal program. You must press the
SatSeconds Alarm Limit button to initiate printing. Refer to Trend
Data Printout on page 57.
#2: Trend Clear
Clears the existing trend data. To activate, press the SatSeconds
Alarm Limit
button.
#3: Not Used