N-560 Operator’s Manual 29
Setting Alarm Silence Duration
Discussion
WARNING: Do not silence an audible alarm or decrease its
volume if patient safety could be compromised.
Alarms can be silenced for a preset period called alarm silence
duration. To view the current setting, press and hold the
Alarm
Silence
button until 30 SEC, 60 SEC, 120 SEC, or OFF is displayed.
To adjust the settings, press and hold the
Alarm Silence button until
30 SEC, 60 SEC, 120 SEC, or OFF is di
splayed. Continue holding the
Alarm Silence button, and use the Adjust Up button or Adjust Down
button to increase or decrease the value. Possible values are 30, 60,
90, or 120 seconds, or OFF. The OFF selection is discussed under
Disabling Audible Alarms on page 31.
If the
Alarm Silence button is pressed during the alarm silence
duration, the alarm silence duration is ended and the audible alarms
are re-enabled.
Visual indications of an alarm condition cannot be turned off. For
exam
ple, if the %SpO
2 lower alarm limit is exceeded, the alarm can
be silenced for the alarm silence duration, but the %SpO
2 value will
continue to flash.
If the alarm condition is still present when the alarm silence duration
has elapsed, the alarm will sound.
T
he power-on default setting for audible alarm silence duration is set
at the factory to 60 seco
nds. The default setting can be adjusted by
service personnel as described in the N-560 service manual.