Product Manual

Specifications
110
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHZ to
80 MHz
3 Vrms Portable and mobile RF
communications
equipment should be
used no closer to any part
of the N-560, including
cables, than the
recommended separation
distance calculated from
the equation applicable
to the frequency of the
transmitter.
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
Recommended separation distance:
where P is the maximum output power
rating of the transmitter in watts (W)
accord
ing to the transmitter
manufacturer and d is the
recommended separation distance in
meters (m). Field strengths from fixed
RF transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment marked with this
symbol:
Note 1: At 80 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
re
flection from structures, objects and people.
Table 9: Electromagnetic Immunity (Cont.)
Th
e N-560 is intended for use in the electromagnetic environment
specified below. The customer or user of the N-560 should assure that it is
used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment — Guidance