Product Manual

Specifications
108
RF emissions
CISPR 11
Class B The N-560 is suitable for use in
establishments, including
diagnostic establishments and
those directly connected to the
public low-voltage power supply
network that supplies buildings
used for domestic purposes.
Harmonic
Emissions
IEC 61000-3-2
Class D
Voltage
f
luctuations/
flicker emissions
IEC 61000-3-3
Complies
Table 9: Electromagnetic Immunity
The N-560 is intended for use in the electromagnetic environment
specifie
d below. The customer or user of the N-560 should assure that it is
used in such an environment.
Immunity Test
IEC 60601
T
est Level
Compliance
Level
Electromagnetic
Environment — Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV
contact
±8 kV air
Complies Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/
burst
IEC 61000-4-4
±2 kV power
supply lines
±1 kV for
in
put/output
lines
Complies Mains power should be
th
at of a typical
commercial or hospital
environment.
Table 8: Electromagnetic Emissions
The N-5
60 is intended for use in the electromagnetic environment
specified below. The customer or the user of the N-560 should assure
that it is used in such an environment.
Emission Test Compliance
Electromagnetic Environment —
Guidance