Product Manual

Specifications
102
Environmental Conditions
Operation
Temperature 50 ºF to 113 ºF (10 ºC to 45 ºC)
Altitude/Barometric Pressure
-390 m to 3,012 m
(
-1,280 ft. to 9,882 ft.)
70 kPa to 106 kPa
(20.6 in. Hg to 31.3 in. Hg)
Relative Humidity
15% to 95% non
-condensing to be
compliant with IEC 60601
-1, sub-clause
44.5
Transport and Storage (not in shipping container)
Temperature
-4 ºF to 140 ºF (-20 ºC to 60 ºC)
Altitude/Barometric Pressure
-390 m to 5,574 m
(
-1,280 ft. to 18,288 ft.)
50 kPa to 106 kPa
(14 in. Hg to 31.3 in. Hg)
Relative Humidity
15% to 95% non
-condensing over
temperature range of
-4 ºF to 140 ºF
(
-20 ºC to 60 ºC)
Transport and Storage (in shipping container)
Temperature
-4 ºF to 158 ºF (-20 ºC to 70 ºC)
Altitude/Barometric Pressure
-390 m to 5,574 m
(
-1,280 ft. to 18,288 ft.)
50 kPa to 106 kPa
(14 in. Hg to 31.3 in. Hg)
Relative Humidity
15% to 95% non
-condensing
Sensor Power Dissipation
Sensor Dissipation
OXIMAX MAX-N 52.5 mW
O
XIMAX MAX-I 52.5 mW