Product Manual

Specifications
N-560 Operator’s Manual 99
Invalid Button Press Volume level Not changeable
Pitch (±30 Hz) 180 Hz
Pulse width (±20 ms) 70 ms
Pulse repetition interval
(±20 ms) (double burst)
N/A
Repetitions 1
Low Priority Alarm Volume level Adjustable alarm
volume
Pitch (±30 Hz) 500 Hz
Pulse width (±20 ms) 200 ms
Pulse repetition interval
(±20 ms) (double burst)
15000 ms
Repetitions Continually
Medium Priority Alarm Volume level Adjustable alarm
volume
Pitch (±30 Hz) 752 Hz
Pulse width (±20 ms) Nellcor = N/A
IAW IEC 60601-1-8
= 160 ms
IAW EN 475 =
200 ms
Pulse repetition interval
(±20 ms) (double burst)
Nellcor = N/A
IAW IEC 60601-1-8
= 7600 ms
IAW EN 475 =
20000 ms
Repetitions Continually
Table 6: Tone Definition (Cont.)
Audible Indicator Parameter Value