Product Manual
Specifications
98
Table 6: Tone Definition
Audible Indicator Parameter Value
Alarm Silence
Reminder
Volume level Not changeable
Pitch (±30 Hz) 784 Hz
Pulse width (±20 ms) 150 ms
Pulse repetition interval
(±20 ms) (double burst)
150 ms
Repetitions 3
Confirmation of
Button Pressed
Volume level Not changeable
Pitch (±30 Hz) 784 Hz
Pulse width (±20 ms) 150 ms
Pulse repetition interval
(±20 ms) (double burst)
150 ms
Repetitions 3
High Priority Alarm Volume level Adjustable alarm
volume
Pitch (±30 Hz) 932 Hz
Pulse width (±20 ms) Nellcor = 255 ms
IAW IEC 60601-1-8
= 120 ms
IAW EN 475 =
150 ms
Pulse repetition interval
(±20 ms) (double burst)
Nellcor = 320 ms
IAW IEC 60601-1-8
= 6940 ms
IAW EN 475 =
7500 ms
Repetitions Continually