Item Brochure

Adult specifications are shown for the OxiMax MAX-A and MAX-N
sensors with the OxiMax N-560 pulse oximeter. Neonate specifica-
tions are shown for MAX-N sensors with the OxiMax N-560 pulse
oximeter. Saturation accuracy will vary by sensor type. Refer to the
Nellcor
Sensor Accuracy Grid Card for details. The low perfusion
specifications apply to monitor performance. Refer to the OxiMax
N-560 pulse oximeter operator’s manual for complete description,
instructions, warnings, cautions and specifications. Specifications
are subject to change without notice.
NELLCOR
OXIMAX N-560 PULSE
OXIMETER FEATURES
Monitors neonates through adults.
Simple, intuitive operation and space-saving
design.
The Covidien distinctive digital signal processing
technology enhances performance during low
perfusion, patient motion and other forms of
signal interference.
Variable pitch beep tone enables clinicians to
hear point-by-point changes in SpO
2
.
Display changes from green to red during alarms.
Eight-hour battery life.
24-hour trend memory.
Optional paging system supplements bedside and
central station alarms to reduce response time.
Integrated Score analysis functionality stream
lines in-hospital sleep screening.
Patient trend data can be stored on a USB drive
for archive and analysis.
TEAM UP WITH TELEMETRY
To enhance patient safety monitoring on your
general care floor, you can team OxiMax N-560
pulse oximeters at the bedside with the Covidien
cost-effective remote oximetry alarm systems or
connect them to nurse call systems.
FEATURES AND SPECIFICATIONS
PERFORMANCE
Measurement range
SpO
2
: 1% to 100%
Pulse rate: 20 to 250 beats per minute (bpm)
Pulse amplitude: 0.03% to 20%
ACCURACY
Saturation (% SpO
2
± 1 SD)
Adult: 70% to 100% ± 2 digits
60% to 80% ± 3 digits
Neonate: 70% to 100% ± 2 digits
60% to 80% ± 4 digits
Low perfusion: 70% to 100% ± 2 digits
Pulse rate 20 to 250 bpm ± 3 digits
Low perfusion: 20 to 250 bpm ± 3 digits
ELECTRICAL
Instrument
Power requirements: 100 to 240 VAC, 25 to 39 VA
Fuse rating: Qty 2, 2A, 250 volts, slow-blow
IEC (5x20 mm)
Battery
Type: NiMH
Battery capacity: Minimum of 8 hours using new, fully
charged battery with no alarms, no
serial data
ENVIRONMENTAL
Operating temperature
Instrument: 10°C to 45°C
Transport/Storage Temperature (in shipping carton):
-20°C to 70°C
Operating Humidity: 15% to 95% noncondensing
Operating Altitude: -1,280 ft to 18,288 ft
PHYSICAL CHARACTERISTICS
Weight 3.07 lb (1.39 kg)
Size
2.95 H x 9.06 W x 5.04 D (inches)
(7.5 cm x 23.0 cm x 12.8 cm)
Equipment compliance
IEC 60601-1 IPX1
CSA 601.1 IS0 9919:2005
UL 2601-1 EN/IEC 60601-1-2 (2nd ed.)
Emissions compliance
EN55011, Group 1, Class B
Output
EIA-232 digital output from serial port
OxiNet Communications Output Protocol
Nurse call capability
Display/indicators
High quality LED display changes from green to red during
alarm conditions.
Pulse amplitude indicator (10-segment LED).
Visual indicators: Pulse search, audible alarms silenced or
off, interference indicator, battery charging, data in sensor
and Nellcor
SatSeconds alarm management clock
Alarms
Audible and visual alarms for high/low saturation and
pulse rate.
SatSeconds Alarm Management settings:
10, 25, 50 and 100, or OFF
Audible and visual warning indicators for low battery and
sensor off.
Audible sensor disconnect alarm.
Optional accessories
GCX mounting adapter
Pole mount clamp
Interface cables
Visual alarm indicator antenna
NELLCOR
OXIMAX SENSOR ADVANTAGES
As part of the Covidien Nellcor
OxiMax pulse oximetry system, the OxiMax N-560 pulse oximeter works with the full line of
OxiMax sensors, which help expand your patient monitoring options.
COVIDIEN, COVIDIEN with logo, Covidien logo and
positive results for life
are
U.S. and internationally registered trademarks of Covidien AG. Other brands are
trademarks of a Covidien company. ©2006, 2011 Covidien. All rights reserved.
11-PM-0264 MN04806
 G A
B, CO

--
..
• OxiMaxSpO
2
forehead sensor offers improved performance
when patients have low perfusion.
• OxiMaxnonadhesivesensorsprovideagentle,single-
patient-use alternative to adhesive sensors for patients with
fragile skin.