Technical data

82
HB EML3020 A-(en)LevelWave LR01
If an "Out of specification" or a "Maintenance" status symbol is shown, refer to menu item 2.2.2
DIAGNOSTIC (Configuration mode / Supervisor menu) for more data.
For data on errors, error records and error codes, refer to Error handling on page 83.
NE 107
symbol
shown
NE 107
Status
Description Status
marker
shown
Error code (Type) Possible errors
Failure The device does not
operate correctly. The fault
message stays on. The
user cannot remove the
"Failure" message from the
Normal mode screen.
1 ERR 101 (Error) Current Output Drift
3 ERR 102 (Error) Temperature Out of Range
1 ERR 103 (Error) Converter EEPROM
1 ERR 103 (Error) Converter RAM
1 ERR 103 (Error) Converter ROM
1 ERR 104 (Error) Converter Voltage
1 ERR 200 (Error) Internal Communication
2, 4 ERR 201 (Error) Overfill
2, 4 ERR 203 (Error) Peak Lost (Level Lost)
3 ERR 204 (Error) Sensor no Signal
1 ERR 206 (Error) Sensor Microwave
1 ERR 207 (Error) Sensor EEPROM
1 ERR 207 (Error) Sensor RAM
1 ERR 207 (Error) Sensor ROM
1 ERR 208 (Error) Sensor Voltage
1 ERR 210 (Error) Sensor Not compatible
Function
check
The device operates
correctly, but the measured
value is incorrect. This fault
message is only
temporary. This symbol is
shown when the user
configures the device with
the DTM or a HART®
Communicator.
- - -
Out of
specification
It is possible that the
measured value is
unstable if the operating
conditions do not agree
with the device
specification.
4 (Warning) Peak Lost
4 (Warning) Overfill
3 (Warning) Temperature out of range
Maintenance The device does not
operate correctly because
of bad environmental
conditions (e.g. build-up on
the antenna). The
measured value is correct,
but maintenance is
neccessary a short time
after this symbol is shown.
5 (Warning) Empty Spectrum Invalid
4 (Warning) Signal Weak
4 (Warning) Signal Strong
4 (Warning) Bad Measurement Quality
3 (Warning) Temperature <-35°C/ -31°F
3 (Warning) Temperature >+75°C/ +167°F