Technical data
56
HB EML3020 A-(en)LevelWave LR01
5.3.2 Functions of keypad buttons
For data on keypad functions, refer to Normal mode on page 58.
5.4 Remote communication with PACTware™
PAC Twa re ™ displays measurement information clearly on a computer (PC) and lets you
configure the device from a remote location. It is an Open Source, open configuration software for
all field devices. It uses Field Device Tool (FDT) technology. FDT is a communication standard for
sending information between the system and the field device. Field devices are easily integrated.
Installation is supported by a user-friendly Wizard.
Figure 5-2: Local display screen layout in configuration mode
1 Function name
2 Configuration mode symbol
3 Menu number
Keypad button Function
[Right]
Normal mode: Enter Information menu (Enter Configuration mode)
Configuration mode: Move cursor to the right
[Return / Escape]
Normal mode: Change units (m, cm, mm, in, ft)
Configuration mode: Exit
[Down]
Normal mode: Change measurement type (distance, level , output (%),
output (mA), conversion, ullage conversion, reflection) 1
Configuration mode: Decrease value or change parameter
[Up]
Normal mode: Change measurement type (distance, level , output (%),
output (mA), conversion, ullage conversion, reflection) 1
Configuration mode: Increase value or change parameter
1 If you have made a strapping table in menu item 2.8.1 INPUT TABLE for volume or mass measurement, "Conversion"
and "Ullage Conv." will be shown in the list of measurement types