H ygrocontrol GmbH HYGROCONTROL TYPE 72 MANUAL Version 08 / 05 Feuchte- und Temperatur-Meßtechnik Fax.: 06181-927930 Moselstraße 2 B, D-63452 Hanau, Tel.: 06181-92790 Internet: www.hygrocontrol.de e-mail: info@hygrocontrol.
Series 72 Manual page TABLE OF CONTENTS: 1. Fundamental Features of Series 72 1 2. Preparations before Operating 1 2.1. 2.2. 2.3. Unpacking and Checking Setup for Operation Instructions for Dispatch 1 2 2 3. Dialog orientated Software 2 3.1. 3.2. Program Structure Pushbutton Operation 2 2 4. Measurement Mode "RUN" 3 4.1. 4.2. Conditions after Power Connection Values in the Display 3 3 5. Software Modul "LIST" 3 5.1. 5.2. 5.3. 5.4. 5.5.
Series 72 Manual page 7.6. 7.7. Calibration of Temperature ("TEMPER") Entering the Pressure Factor ("PRESS") 10 10 8. Configuration of Series 72 Transmitters 11 8.1. 8.2. 8.3. 8.4. 8.5. 8.6. 8.6.1. 8.6.2. 8.6.3. 8.6.4. 8.6.5. 8.7. 8.7.1. 8.7.2.
Series 72 Manual 12. Diagrams of Program structure Main Programs Program "LIST" Submenu "LIST - MINMAX" Submenu "LIST - CALDAT" Submenu "LIST - CONDAT" Program "CALIB" Program "CONFIG" Submenu "CONFIG - DIM" Submenu "CONFIG - SERIEL" Submenu "CONFIG - RELAIS" 13.
Series 72 Manual page 1 1. Fundamental Features of Series 72 Series 72 Humidity and Temperature Meters detect relative Humidity and Temperature. For this purpose the probes - either fixed mounted or interchangeable ones - are equipped with a capacity type humidity sensor of very small mass. The temperature detection is done by a RTD of type Pt-1000, which is placed very close to the humidity sensor to avoid differences in temperature between the two sensors.
Series 72 Manual page 2 2.2. Setup for Operation The instrument is to be connected to main powers either by a connector or by using the terminals inside the housing. The correct voltage is marked on the type label you find on one side of the housing. Do not use voltages other than the marked one! Severe damage would be the result of applying the wrong voltage. If a type 72B. instrument with screw terminals is used, the cross section of your conductors should meet the dimensions of 0.5 up to 1.
Series 72 Manual page 3 done by turning the four fast mounting screws 90 degrees to the left with a screwdriver. The red coloured buttons are marked on the display - board with: 1) 2) 3) 4) "E" "M" " " " " = enter = mode = encrease = next decimal The two arrows are the only pushbuttons which have an "enhanced speed" function, which operates after pressing them for a time period of more than 1 second. In the "LIST" submenus the " "-button leads to the next, the " "-button to the previous display point.
Series 72 Manual page 4 This is a feature which enables the user, supplier and manufacturer - even after a malfunction - to analyse what has happened with the transmitter.
Series 72 Manual page 5 5.4. Data in "LIST" - "CONDAT" Submenu "CONDAT" displays all data, which describe the actual configuration of the transmitter. After chosing "CONDAT" and pressing "E", you will find the data as follows: 1) lower limit of Humidity range 2) upper limit of Humidity range 3) reaction at sensor break Hum. 4) lower limit of Temp. range 5) upper limit of Temp. range 6) reaction at sensor break Temp. 7) offset analog output Temp. low 8) offset analog output Temp.
Series 72 Manual page 6 F) = sensor oscillator R) = reference oscillator Finally the display shows the actual pressure factor, marked by "PRESS". This factor is used to calculate humidity values at atmospheric pressures, if measurement occurs at other pressure conditions. The calibration data display may be left by pressing "E", the program then jumps back to "CALDAT". 6.Calibration 6.1. General Notes about Calibration Standards There is a lack of international standards for rel.
Series 72 Manual page 7 To assure the easiest handling and highest accuracy, the solutions are enclosed in glass ampoules, which should be opened only before usage. Opening is simplified by the weakened middle part of the ampoules. Once opened the solutions may catch humidity from or give humidity to their environment - so the time of usage is restricted and they should be used for one calibration only. 6.4. Calibration Chamber The calibration chambers we are supplying under No.
Series 72 Manual page 8 Ensure that the calibration chamber faces downwards, to avoid direct contact of the solution with the sensor. Before inserting the bottom into the chamber, unscrew the chamber slightly, to help escaping air, when you close the chamber. Tighten the chamber again after closing! 7. Program "CALIB" 7.1. Password Protection To protect series 72 instruments against unwanted changes of their calibration, this part of the program is protected by a password.
Series 72 Manual page 9 1) A calibration with saturated LiCl-solution of 11.3% will replace the former 10%-value. 2) A calibration with saturated NaCl-solution of 73.6% will not replace the two former values at 65% and 80%. 7.3. Entering "CALIB" From normal measuring mode you enter the main program area by pressing "E". The display now shows "LIST". Pressing "M" now switches over to "CALIB" and the first line of the display shows "0000" with a flashing first digit.
Series 72 Manual page 10 7.5. Submenu "DELETE" The submenu "DELETE" allows you to cancel an old calibration totally, before recalibrating a probe. This procedure must be done, when tests show a deviation of standards of more than 5% rel. Humidity. In this case, check the sensor by visual inspection. If no mechanical defects occured, and no layer of dust or other contamination is found, a new calibration is required. To erase all calibration points hit "E" when "DELETE" is in the display.
Series 72 Manual page 11 8. Configuration of Series 72 Transmitters 8.1. Password Protection As in "CALIB", the program "CONFIG" has password protection to avoid unwanted changes of configuration. The same 4-digit number is used as in "CALIB". 8.2. Possible Configurations The digital electronics of type 72 instruments allow the user to fully customize the display and outputs. We tried to do this in a practical manner to adapt the instrument to a variety of applications.
Series 72 Manual page 12 8.6. Submenu "ANALOG" This part of the program is used to define the ranges of temperature and humidity, which correspond to the analog outputs. Furthermore this menu may adjust offsets in the analog outputs. This is helpful to compensate for any discrepancies between the series 72 instrument and any other device coupled to the output. 8.6.1.
Series 72 Manual page 13 8.6.5. Limits of Temperature Range In the same way you may now change temperature limits and reaction of analog output, if a break in the sensor line of Pt-1000 occurs. The display will now show "T out" instead of "H out" or "T err" instead of "H err", but the change of values is the same as points 7.6.2, 7.6.3 and 7.6.4. If these parts of the program are done, the "RANGE" menu ends.
Series 72 Manual page 14 9.1.1. Protocol 1 Transmission of datas will occure every second. Each string consists of 16 ASCII-codes, configurated as follws: STX DIM SPACE HUMIDITY SPACE TEMPERATURE CR LF BYTES 1 1 1 5 1 5 1 1 DECIMAL 2 s.b. 32 s.b. 32 13 13 10 HEXADECIMAL 2 s.b. 20 s.b. 20 s.b. OD OA DIM: The dimension will be sent as ASCII-Code.
Series 72 Manual page 15 00809 is to read 80.9 %RH (or other dimension) 9.2. MODUS 2: Software Handshake, controlled by Host It is necessary that: Signal CTS is connected to +12V. (Bridge RTS-CTS) 9.2.1. Control of stand-alone Transmitter (XON/XOFF-Handshake) For this, the address of the transmitter must be "0" The following ASCII-codes will cause actions: XON: XOFF: CAN: P: (17dez,11hex) String as in Mode 1 will be send. (19dez,13hex) ends data transfer after next Byte and deactivates digital output.
Series 72 Manual O: page 16 (79dez,4Fhex) Output: opens the ring, sends a string of datas (with address) according to protocol 2 (s.b.). closes the ring thereafter and so enables further transmissions. 9.2.3. Protocol 2: A string of 19-ASCII codes is transmitted with the following content: STX ADR SPACE DIM SPACE HUMI. SPACE TEMP. CR LF BYTES 1 2 1 1 1 5 1 5 1 1 DECIMAL 2 s.b. 32 s.b. 32 s.b. 32 s.b. 13 10 HEXADECIMAL 2 s.b. 20 s.b. 20 s.b. 20 s.b.
Series 72 Manual page 17 Humidity: Again this value is sent with resolution of 1/10 %RH (or any other dimension) without decimal point as ASCII-code. The highest decimal is sent first. Dewpoints may have negative values which are transmitted as their dualcomplement (16-Bit), without sign, in the same way as negative temperatures. Example: 00809 is to read 80.9 %RH (or other dimension) 9.3.
Series 72 Manual page 18 message (flashing) will be displayed in the lower line. This message shows as "err...." where "..." is a three digit number. This number gives information about the nature of the malfunction detected: 001 = No Humidity frequency = Defect of Humidity sensor. 002 = No reference frequency = Mostly caused by cable break. 004 = Temperature < -50,5 °C = Short circuited Pt-1000 005 = 001 and 004 are both detected. 006 = 002 and 004 are both detected.
Series 72 Manual page 19 11.1 Direct Contact with Liquids Avoid in any case direct contact between the humidity sensor and any liquid. The sensor may only detect the humidity in the atmosphere over the surface of a liquid or a solid material. Even if the sensor will not be changed in his characteristic by direct contact with water and will measure correct values after being dryed, you should avoid immersion or condensation.
Series 72 Manual page 20 11.4 Protection of Transmitter Electronics To give highest protection against the surrounding atmosphere, our transmitter housings are according to IP 65. Avoid atmospheres of more then 80 %RH, as condensation could occur with rapid temperature changes. 11.5 Influence of Temperature on Humidity To detect rel.
Series 72 Manual page 21 PSF = Saturated Vapor Pressure over Chilled Water PSE = Saturated Vapor Pressure over Ice M = Rel. Humidity displayed by HYGROMESS Instruments. 11.7 Measurements under Pressure or Vaccuum All probes can be used at pressures of 0,9 to 1,3 bar. Only special probes with a glas feed-through (which predicts SS - 316 tube material) may be used under pressures of 0,03 to 30 bar. As rel. Humidity is directly proportional to the pressure (f.E.
Series 72 Manual page M 1 MAIN PROGRAMS MEASURING E LIST E M 0000 CALIB E TO MENU LIST, SEE PAGE M2 M E TO MENU CALIB, SEE PAGE M6 M 0000 CONFIG RUN E TO MENU CONFIG, SEE PAGE M7 USED SYMBOLS: ABCDEF PROGRAM 0000 ABCDEF E 0000 ABCDEF PROGRAM: ACCESS ONLY BY KEYCODE PUSHBUTTONS TO USE: NEXT DECIMAL ABCDEF ACTION TO COME INCREASE VALUE M RETURN WITHOUT NEW VALUE E PROCEED WITH NEW VALUE M
Series 72 Manual page M 2 PROGRAM “LIST” (DISPLAY OF DATA IN MEMORY) LIST E M MINMAX M CONDAT E E M CALDAT M RETURN E E SEE PAGE SEE PAGE SEE PAGE M3 M4 M5 To next higher Menue Level M RUN E Back to Measurement Mode
Series 72 Manual page M 3 SUBMENU LIST - “MINMAX” (DISPLAY OF MIN. AND MAX. VALUES) 20.0°C T - min MINIMUM OF TEMPERATURE VALID OPERATIONS IN THIS SUBMENUE ARE: 90.1°C T - max MAXIMUM OF TEMPERATURE PREVIOUS DISPLAY 10.1% H - min MINIMUM OF HUMIDITY NEXT DISPLAY 95.1% H - max MAXIMUM OF HUMIDITY 20.1°C LAST VALUE OF TEMPERATURE T-1 20.3°C T-2 21.2°C T-3 79.6% H-1 79.7% H-2 79.
Series 72 Manual page M 4 SUBMENU LIST - “CALDAT” (DISPLAY OF CALIBRATION DATA) 7 ANZAHL NUMBER OF POINTS CALIBRATED 0.1% a 12464 CALIBRATION POINT “a” PREVIOUS DISPLAY AND SO ON UNTIL NEXT DISPLAY 95.1% k 13578 -23.4 1.07423 00.0% 2.14523 80.0% CALIBRATION POINT “K” E OUT OF SUBMENUE DATA OF STRAIGHT LINE CALIBRATION M *) MARKED VALUES MAY BE HOLD PRESSING THIS BUTTON FIRST CALIBRATION AT FACTORY #) 3.
Series 72 Manual page M 5 SUBMENU LIST - “CONDAT” (DISPLAY OF CONFIGURATION DATA) 7 H out LOWER LIMIT OF HUMIDITY OUTPUT VALID OPERATIONS IN THIS SUBMENUE ARE: 0.1% H out UPPER LIMIT OF HUMIDITY OUTPUT PREVIOUS DISPLAY 0.1% H err CONDITIONS AT SENSOR BREAK NEXT DISPLAY 95.1% LOWER LIMIT OF TEMP. OUTPUT T out -23.4 T out 00.0% E OUT OF SUBMENUE UPPER LIMIT OF TEMP. OUTPUT 4800 CONDITIONS AT SENSOR BREAK RS 232 T set ANALOG OUTPUT AT LOW TEMP.
Series 72 Manual page M 6 PROGRAM “CALIB” (CALIBRATION OF HUMIDITY AND TEMPERATURE) M M CHANGE M DELETE E TEMPER E M E E All values deleted in the memory M XX % M 0% M 10 % E E E CHECK IF VALUE IS STABLE! YOU WANT TO KEEP IN MEMORY? N 33,5% Y M VALUE NOT STORED IN MEMORY! AND SO ON UNTIL M E E M To next higher Menue Level 100 % M Back to Measurement Mode M OTHER RETURN E E E CHOOSE REL.
Series 72 Manual page M 7 PROGRAM “CONFIG” (CONFIGURATION OF RANGES AND OUTPUTS) M M DIM ANALOG E RANGE RELAIS*) E M M OFFSET E M RETURN M H SET M M H set M M H out H out E E Enter Value E E Enter E Back to Measurement Mode M T out E Enter Value E RUN RETURN M H err Back to Measurement Mode E E E M To next higher Menue Level 2000 T set 2000 To next higher Menue Level E E E E E RETURN 000 T set H set RUN Back to Measurement Mode E 000 M E T SET E E
Series 72 Manual page M 8 SUBMENU “CONFIG - DIM” (CHOOSE DIMENSIONS OF HUMIDITY + TEMP.) E M HUMI M E Choose Humidity Dimension M TEMP.
Series 72 Manual page M 9 SUBMENU “CONFIG - RELAIS” (ENTER RELAIS FUNCTIONS) M M REL 1 E REL 2 M REL 3 E E M M REL 4 RETURN E E M ENTER SWITCH POINT E E RELAY ON ABOVE OR BELOW ENTER HYSTERESIS E ENTER HYSTERESIS E M ENTER SWITCH POINT RELAY ON ABOVE OR BELOW E M ENTER HYSTERESIS E M ENTER SWITCH POINT E RELAY ON ABOVE OR BELOW E M E E RELAY ON ABOVE OR BELOW E Back to Measurement Mode E M ENTER SWITCH POINT E CHOOSE HUMIDITY OR TEMP. CHOOSE HUMIDITY OR TEMP.
Series 72 Manual page C 1 TYPE 73 (19”): 64- OR 32-PIN CONNECTORS PIN 1A 1C 2A 2C 3A 3C 4A 4C 5A 5C 6A 6C 7A 7C 8A 8C 9A 9C 10A 10C 11A 11C 12A 12C 13A 13C 14A 14C 15A 15C 16A 16C 17A 17C 18A 18C 19A 19C 20A 20C 21A 21C TRANSMITTER I²C SDA I²C SCL +5 V GND TMESS0 IKONST CLK SDO FREF FHUMI TMESS3 TMESS4 U OSZ GND U FAN SDI -5 V +15 V +5 V +5 V GND GND ANALOG HUMI ANALOG GND ANALOG TEMP ANALOG GND RxD TxD RTS CTS GND GND TAST ENTER TAST COUNT TAST SHIFT TAST MODE ANALOG IN 1 ANALOG IN 2 ANALOG IN 3 ANALOG
Series 72 Manual page C 2 TYPE 73 (19”): 64- OR 32-PIN CONNECTORS PIN 22A 22C 23A 23C 24A 24C 25A 25C 26A 26C 27A 27C 28A 28C 29A 29C 30A 30C 31A 31C 32A 32C TRANSMITTER INT 3 INT 1 N PE PE L1 LEGEND: () () = = = = = = = = = + = = + + = = = + = = MULTIPLEXER CLK SDO FREF FHUMI TMESS3 TMESS4 U OSZ GND U FAN SDI TMESS0 IKONST CLK SDO FREF FHUMI TMESS3 TMESS4 U OSZ GND U FAN SDI 3 3 3 3 3 3 3 3 (3) 3 4 4 4 4 4 4 4 4 4 4 (4) 4 / : I²C-BUS LINES < : PROBE INPUT TO TRANSMITTER * : ANALOG OUTPUT # : DIGI
Series 72 Manual page C 3 CONNECTIONS / DIMENSIONS OF TYPE “A” (1) (2) (1)= POWER (2) = SIGNAL-OUTPUT (3) = SENSOR-INPUT 29 (1) + (2) = HAN 3A (3) = BINDER 423 50 PIN CONNECTION ANALOG OUTPUT (2): 82 70 PIN 1 = - HUMIDITY PIN 2 = + HUMIDITY PIN 3 = + TEMPER. PIN 4 = - TEMPER. (3) PIN CONNECTIONS AT FIXED CABLE VERSIONS: BOTTOM VIEW WHITE YELLOW GREEN BROWN 2 1 3 3 2 1 55 = - HUMIDITY = + HUMIDITY = + TEMPER. = - TEMPER.
Series 72 Manual page C 4 CONNECTIONS / DIMENSIONS OF TYPE “B” PIN CONNECTION ANALOG OUTPUT: 4-POL. CONNECTOR: PIN 1 = - HUMIDITY PIN 2 = + HUMIDITY PIN 3 = + TEMPER. PIN 4 = - TEMPER. PIN CONNECTIONS AT FIXED CABLE VERSIONS: WHITE YELLOW GREEN BROWN = - HUMIDITY = + HUMIDITY = + TEMPER. = - TEMPER. SCREW TERMINALS: B 7-1 B 7-2 B 7-3 B 7-4 = - HUMIDITY = + HUMIDITY = + TEMPER. = - TEMPER.
Series 72 Manual page C 5 CONNECTIONS HYGROCONTROL 72B - DIGITAL CONNECTION OF DIGITAL OUTPUTS RS-232C, RS-232.TTL UND RS-485 AT DIFFERENT TRANSM TERMINALS : 6-pol.
Series 72 Manual page C 6 CONNECTIONS OF SPECIAL VERSIONS A) 7 - POL.-CONNECTOR TYPE C 16-1 3 PIN NUMBER 1 2 3 4 5 6 4 5 2 1 6 CONNECTED - HUMIDITY + HUMIDITY + TEMPER. L1 (POWER) N1 (POWER) - TEMPER. PE (POWER) B) 2 x 4 - POL.
Series 72 Manual page C 7 ACCU - VERSIONS ACCU red green Switch Control Elements Switch on off LED red off blinking LED green on off Power switch on in Backplane off Main Power Accu Operation Transmitter on Transmitter off Accu empty Accu is good for 1h Power Supply o.k. Power is off Power on Power off The Accu has a Voltage of 12 V (2x6 V Lead-Gel) with a capacity of 1,1 Ah. This lasts for a running time of at least 5 hours for the transmitter.