Specifications
Luminescence Decay Measurement in the Microsecond Range 23
Fig. 29: Trace parameters and display parameters recommended for fluorescence decay measurement
Fig. 30: Decay curves recorded with the setup parameters shown in Fig. 27 through Fig. 29
Luminescence Decay Measurement in the Microsecond Range
A suitable setup for luminescence decay measurements in the µs range is shown in Fig. 31.
For extremely long decay times the laser pulse repetition rate must be reduced to a few 10 kHz
or less. Using ps pulses at a repetition rate this low would result in an extremely low average
excitation power. It is therefore better to use excitation pulses with a duration of a few 100 ns
to several µs. The bh BDL-SMC diode laser is therefore operated in the CW mode [5]. The
laser is switched on and off periodically via its on/off input. To generate pulses of appropriate
duration suitable on/off signal is generated by a bh DDG-200 pulse generator module. The
pulse width and repetition rate of the excitation pulses is defined by the DDG-200 setup pa-
rameters.
SYNC
Trigger Out
+12V
Detector
DPC-230 module
PMC Power supply and
DCC-100 Detector Controller
bh PMC-100
Photon
ps Diode laser
A-PPI-D Pulse
Overload Shutdown
Filter
Sample
Laser Switch Box
bh BDL-SMC
Inverter
Delay
DDG-200 Pulse Generator
Module
or other pulse generator
Power
on/off
A-PPI-D
pulse
inverter
CW
Fig. 31: Luminescence decay measurement in the microsecond range










