User Manual

8
XD730.400 Piezo Objective Scanner User Manual
3. After checking for no abnormalities, input voltage values according to requirements
(keyboard or analog control).
4. It could work normally within the rated voltage range of piezo ceramic (0~120V, max not
exceeding 150V);
5. Please set the piezo controller to zero after used.
Please do not touch the product during the operation, avoid affecting the piezo scanner
performance or even damage the piezo scanner.
The resolution of piezo ceramic is not limited. It depends on the performance of piezo
controller. In order to better achieve the motion accuracy, it is recommended to use
CoreMorrow piezo controller. For the operation of piezo controller, please refer to piezo
controller usual manual.
Note:
Please do not remove the grounding protection during operation. If it needs to be
temporarily removed, reconnect the grounding protection before starting again.
Drive voltage should not exceed 150V.
If abnormal noise or oscillation occurs during operating, immediately turn off the
power check parameter setting.
6.1 Transportation
Piezo stage is packed in carton. The transportation is carried out under the condition of
packaging. This product can be transported by various means of transportation under
normal conditions. During transportation, it should avoid direct rain and snow, avoid
contact with corrosive substances, and avoid strong collision, try to avoid improper
behavior such as squeezing, irregular placement.
6.2 Unpacking inspection
Please check below information:
Before opening the package, please carefully check the packaging for damage,
6. Transportation and inspection