User Manual
6
P76 Piezo Objective Scanner User Manual
P76
Adapter
Lens
Figure2. P76 Installation diagram
Installation steps of P76, as follows:
1. Screw the corresponding objective lens into the P76 objective scanner and tighten;
2. Screw the lens into the adapter and tighten;
3. Fix the installed piezo objective scanner and microscope connection section.
5. Installation and operation
5.1 Installation and connection
Note:
During the installation process, please pay attention to handle, so as not to damage the
P76 piezo objective scanner or scratch and wear it.
When connecting P76, it is required that the weight of the objective lens does not
exceed the maximum allowable load of P76.
Connect the cable to the corresponding interface of the controller according to the
identification on the connector.
Before installation, first check the safety, such as whether the cable is damaged or broken,
and whether the power supply has grounding protection.