User Manual

P73 Piezo Objective Scanner User Manual
8
When connecting the objective lens, it should be noted that the mass of the objective lens
should not exceed the load capacity of P73, and the load should not be too long, so as to
P73 may be damaged by oscillation.
5.3 Operating
Please reconfirm the following information before turning on the power:
The operating environment meets the operating environment requirements of P73.
Fixed screws must be tightened.
After the above information is confirmed, relevant controller operations can be carried out:
1.Please read the user manual of the controller carefully before operation, and then start
the controller.
2.Set the controller' output voltage to zero first.
3.Check no abnormal phenomenon, then input voltage value (via keyboard or analog
control) according to specified input voltage range.
4.Operating normally within the rated voltage range of piezoelectric ceramics (0~120V,
maximum 150V).
5.After use, please carry out voltage reset operation.
Please take light steps when P73 in operation, do not knock the product, so as not to
affect the use of the product or even damage the product.
The resolution of piezo is not limited. It depends on the performance of piezo Controller. In
order to achieve better motion precision of P73, it is recommended to use our company's
supporting controller. Please refer to the user manual of the controller for the specific
operation to ensure correct operation before relevant operation.
Notice:
The driving voltage shall not exceed the voltage range of P73.
If any abnormal noise or oscillation occurs during use, turn off the power immediately
and check the parameter setting.