Service manual
GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION , REVISION
3-34 Section 3-10 - Loading Base System Software
3-10-3 Base Load Software Load Procedure For BEP Ver 2.0
NOTE: Disconnect the system from the network and remove all transducers, all External Drives (USB
MOD, Flash Sticks, USB Memory etc.,), Switch Off peripherals like (VCR, VGP) before
installing Base Load.
1.) Insert the disk labeled “Base System Software Load Image as per BEP Type” into the CDROM
drive and switch on the system. The following screen appears, Press any key to Continue.
2.) Select Option 1.
3.) The second choice is to load only the bootable portion of the hard drive partition, the “C”
partition. This is not an option for a new hard drive that has not been previously partitioned for
proper LOGIQ 3 operation.
Note: This selection will leave the remaining drive partitions intact.
Figure 3-17 Base Load Warning Message
Figure 3-18 Base Load Option Message