Service manual

GE MEDICAL SYSTEMS
D
IRECTION 2300749-100, REVISION 2LOGIQ
TM
3 SERVICE MANUAL
Chapter 3 3-15
3-4-5-2 Scanner Shutdown
1.) Switch OFF the Circuit Breaker at the back of the system.
2.) Disconnect the Mains Power Cable if necessary. For example: Relocating the scanner.
Figure 3-11 Circuit Breaker and Power Cable on Back of Scanner
NOTICE
Disconnecting the Mains Power Cable before switching OFF the Circuit Breaker will activate the
uninterruptible power system (UPS) in the backend processor, forcing an ordered shutdown of the
system.
NOTICE
The UPS serves only as a battery backup that allows for a soft shut-down of the scanner to prevent file
corruption. IT DOES NOT REGULATE INCOMING VOLTAGE.
Circuit Breaker
Mains Power Cable
Cable Clip