Service manual

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION , REVISION
3-14 Section 3-4 - Completing the Installation
3-4-4-2 Back-end Processor Boot Up
Press the Standby switch on the Control Panel once.
When the Standby switch on the Control Panel is pressed once, the Power is distributed to the Front-
End, to the Peripheral outlets and to the Back End Processor. The Back End Processor starts and the
software is initiated to start the scanner.
No status messages are displayed during this process.
3-4-5 Power Off/ Shutdown
NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be
able to boot if power is recycled too quickly.
3-4-5-1 Back-end Processor Power Down
1.) Press the Standby switch on the Control Panel once to display the SYSTEM - EXIT menu.
2.) Select Shutdown from the SYSTEM - EXIT menu.
Figure 3-9 Standby Switch for Back-end Processor Boot Up
Figure 3-10 System Exit Menu for Back-end Processor Power Down
3
Standby Switch