3-12 Section 3-4 - Completing the Installation
3-4-1-2 Weight with Monitor and without Peripherals
The Weight of LOGIQ™ 3 with monitor, without peripherals
3-4-1-3 Acoustic Noise Output:
Less than 70dB (A) according to DIN 45635 - 19 - 01 - KL2.
3-4-2 Electrical Specifications
Electrical conduit, junction boxes, outlets, circuit breakers, and switches should be in place before
installing the LOGIQ™ 3 console
3-4-3 Probe (Transducer) Connection
1.) Connect a transducer to the upper transducer receptacle as follows:
A.) Ensure that the transducer twist lock lever points towards the 9 o'clock position.
B.) Insert the transducer connector on the receptacle guide pin until it touches the receptacle
mating surface.
C.) Twist the transducer twist lock lever to the 2 o'clock position to lock it in place. Twist the
lever to the 9 o'clock position to disconnect the transducer.
NOTE: It is not necessary to turn OFF power to connect or disconnect a transducer.
Connect the main power cable to a hospital grade power receptacle with the proper rated voltage
checked during pre installation. Never use a three-to-two prong adapter; this defeats the safety ground.
3-4-4 Power On / Boot Up
NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be
able to boot if power is recycled too quickly.
3-4-4-1 Scanner Power On
1.) Connect the Power Cable to the back of the system.
2.) Ensure the Cable Clip slips securely over the shoulders on the molded plug.
3.) Connect the Main Power Cable to a hospital grade power receptacle with the proper rated voltage.
Never use an adapter that would defeat the safety ground.
4.) Switch ON the Circuit Breaker at the rear of the system.
Table 3-5 Weight of LOGIQ™ 3
Model Weight [Kg] Weight [lbs]
LOGIQ™ 3 155 342
Table 3-6 Electrical Specification for LOGIQ™ 3
System Voltage Current Frequency
1 100 ~ VAC 10A 50-60 Hz
2 115 ~ VAC 8.2A 50-60 Hz
3 230 ~ VAC 4,1A 50-60 HZ