Service manual
GE MEDICAL SYSTEMS
D
IRECTION 2300749-100, REVISION 2LOGIQ
TM
3 SERVICE MANUAL
Chapter 10 10-3
Table 10-2 Customer Care Schedule
Service at Indicated Time Daily Weekly Monthly Annually Notes
Clean Probes •* * or before each use
Clean Probe Holders
•
Clean Air Filter •
more frequently depending on
your environment
Inspect AC Mains Cable • Mobile Unit Check Weekly
Inspect Cables and Connectors •
Clean Console •
Clean Monitor •
Inspect Wheels, Casters, brakes and
Swivel Locks
• Mobile Unit Check Daily
Console Leakage Current Checks
•
also after corrective
maintenance
Peripheral Leakage Current Checks •
also after corrective
maintenance
Surface Probe Leakage Current Checks •
also after corrective
maintenance
Endocavity Probe Leakage
Current Checks
Twice
Annually
Transesphongeal Probe Leakage
Current Checks
As Prescribed
in probe manual
Surgical Probe Leakage
Current Checks
As Prescribed
in probe manual
Measurement Accuracy Checks •
also after corrective
maintenance
Probe/Phantom Checks •
also after corrective
maintenance