8-24 Section 8-3 - Upgrading Application Software
8-11-4 Applications Software Upgrade Procedure (cont’d)
7.) As indicated, press any key to continue. The process is completely automated and provides
you with messages as to its status. R1.1.1 Software provides the message in Figure 8-93.
At this point you just have to wait while the software loads (Approximately 10 minutes).
Near the end of the R1.1.1 software install, the installation script will check to see if Direct CD needs to
be updated. If Direct CD is updated, several more windows will appear. These windows will be handled
automatically.
8.) WWhen it completes loading the system will reboot.
9.) Remove the CD from the drive
If possible, while the system is rebooting remove the CD from the drive. If you didn’t do that don’t worry
you will get a dialog box like the one shown in Figure 8-94.
10.)REMOVE the Applications CD
11.)From the Start Applications dialog box, select START
.
12.)Select CANCEL
.
Figure 8-24 R1.1.1 Software Message
Figure 8-25 Start Software