8-22 Section 8-3 - Upgrading Application Software
Section 8-3
Upgrading Application Software
8-3-1 Manpower
One person, .5 hour + travel
8-3-2 Tools
None
8-3-3 Preparations
Although loading applications software is not designed to write over presets, patient data and exams, it
may be prudent to:
- Back up any customer preset configurations if necessary.
- Export and patient exam files if necessary.
Refer to 8-10-4 "Image Management Guide" on page 8-87 for deatils.
8-3-4 Applications Software Upgrade Procedure
1.) Place the “Application Software” CD-ROM into the CD-ROM drive.
2.) Power Down the scanner.
3.) Power up the scanner. Wait for a dialog box like the one in Figure 8-90.
WARNINGWARNING
While the software install procedure is designed to preserve data, you should save any
patient data, images, system setups to a MOD or hardcopy.
CAUTION
Remember to remove the MOD before continuing with the software install.
Figure 8-21 Start Application Window