Service manual

GE MEDICAL SYSTEMS
D
IRECTION 2300749-100, REVISION 2LOGIQ
TM
3 SERVICE MANUAL
Chapter 8 8-11
Section 8-2
Loading Application Software After a Base Image Load
8-2-1 Manpower
One person, .5 hour + travel
8-2-2 Tools
None
8-2-3 Preparations
Loading applications software is necessary after a Base Image Software Load.
If this is the first time for loading Applicatiopns Software (after a Base Image Load) proceed with the
following process.
If this is NOT the first time for loading Applications software (just an applications update) then proceed
to Section 8-11 "Upgrading Application Software" on page 8-100.
8-2-4 Image Management Guide
Save As to View on any PC - Use this to save images (Dicom or Raw Dicom) in a computer-friendly
format (.avi or .jpeg) so you can view it on any PC.
Move Images to Archive - Use this to take images off your Ultrasound system on to removable media
for long-term archive. This is the way to free up hard disk space, rather than deleting images.
Export/Import Data/Images Between Systems - Use this to copy both patient data and images for
specified patient(s) from one system to another.
8-2-4-1 Save As (Saving Images to CD-ROM to View on Any PC)
To save images to the CD-ROM:
1.) Insert the CD.
2.) Go to Utility --> Connectivity --> Tools. Select the Media as CD Rewritable.
3.) Press Format to format the CD-ROM.
4.) Select the image(s) to be saved to CD-ROM, press Menu, and select Save As.
5.) The SAVE AS menu appears.