Service manual

GE MEDICAL SYSTEMS
D
IRECTION 2300749-100, REVISION 2LOGIQ
TM
3 SERVICE MANUAL
Chapter 8 8-5
8-1-4 BEP Power Supply (FRU no. 703)
2.) Unscrew the seven screws (1-7) and remove the side cover.
3.) Remove the cable connections as shown in the figures.
4.) Remove the BEP Front cover by removing the four indicated screws (8-11).
5.) Remove the cable from the cable clamp (a)
6.) Remove the seven screws, marked (12-18), attaching the power supply to the BEP and the four
screws (19-22) attaching the battery to BEP.
7.) Replace the power supply and the battery.
8-1-4-1 Mounting Procedure
1.) Replace the parts in the order of removal.