Service manual
GE MEDICAL SYSTEMS
D
IRECTION 2300749-100, REVISION 2 LOGIQ
TM
3 SERVICE MANUAL
7-10 Section 7-5 - Common Diagnostics
Section 7-5
Common Diagnostics
7-5-1 Utilities
7-5-2 PC Diagnostics (Non-Interactive Tests)
The non-interactive or interactive PC Diagnostic tests are automatically executed only by clicking the
following menu. The test log, displayed at the status frame, reports all diagnostic test results, the testing
time, and testing date.
7-5-2-1 CPU Tests
7-5-2-2 Hard Drive Tests
7-5-2-3 Memory Tests
7-5-2-4 CD-ROM Drive Tests
7-5-2-5 Video Test
7-5-2-6 USB Test
7-5-2-7 PCI Board Configuration Test
7-5-2-8 Keyboard Test
7-5-2-9 Audio Test
NOTE: Monitor Test DoubleOn Systems with Transmit Board 2318124-2 , the TPG2 Test will fail,
even if the ASIC is working fine. This test should be Neglected on system with the Transmit
board 2318124-2.
7-5-2-10
Table 7-2 Utilities Mode
Items
Descriptions
Disruptive
Enable: The disruptive diagnostic mode is ON and the other PC connected to the scanner via network
can operate the scanner remotely for troubleshooting.
System Shut Down
N/A