Service manual

GE MEDICAL SYSTEMS
D
IRECTION 2300749-100, REVISION 2 LOGIQ
TM
3 SERVICE MANUAL
4-22 Section 4-4 - Functional Checks
4-4-6 Measurement and Multi Image Checks
4-4-6-1 Performing B Mode Measurements Checks
Figure 4-16 Measurement Keys
Table 4-12 Measurement Checks
Step Task Expected Result(s)
1.
Generate the B image. A B image is active on the screen
2.
Press Freeze Freeze the image to measure.
3.
Press MEASURE on the Control Panel to
enter the Assign and Measure modality.
The measurement soft menu for the current
exam category is displayed on the Monitor.
Select the appropriate exam category and
measurement kind.
5.
Trackball the cursor
Move the cursor to the start point of the
measurement
6.
Press Set. Anchor the start point of the measurement.
7.
Trackball the cursor
Move the cursor to the measurement end
point.The current distance value is displayed
in the Measurement result table and is
instantaneously updated when moving the
cursor