Service manual
GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION , REVISION
3-48 Section 3-11 - Loading Application Software
3-11-6 System Setup
To complete the system setup:
1.) Connect the scanner to the network if applicable.
2.) From the Touch Screen on the scanner, select Utility>Connectivity>TCPIP Tab.
3.) Ensure that DHCP is deselected so you can enter information manually.
4.) Input the DICOM information.
* This should be the appropriate DICOM information for the system if it is on the hospital
network. If it is not correct, input the correct data.
5.) After all information is entered select SAVE SETTINGS
.
6.) If the system is NOT on a network you will need to input a set of "dummy" data for proper InSite
and Diagnostic operation. Enter the following data if you are not on a network:
* Computer Name:
* IP Address: 3.192.28.253
* Subnet Mask: 255.255.255.0
7.) After all information is entered select SAVE SETTINGS
.
8.) Reset the VCR by going to Utilities>System>Setup.
9.) Select Sony 9500 VCR and SAVE
10.)Select Exit in the lower left-hand corner of the Touch Screen.
11.)Do an auto-shut down
12.)Press the STANDBY
switch on the Control Panel once to display the SYSTEM - EXIT menu.
13.)Select SHUTDOWN
from the SYSTEM - EXIT menu.
3-11-7 Restore Patient, Report, and User-Defined Presets
1.) Place the Backup/Restore media (MOD or CD-R) in the drive.
2.) Select the UTILITY
tab on the Touch Panel. Select SYSTEM, then BACKUP/RESTORE.
3.) Select everything under Restore by placing a check mark after Patient Archive, Report
Archive, System Configuration, and User Defined Configuration. Then press RESTORE
.
Figure 3-32 System Exit Menu for Back-end Processor Power Down