Service manual
GE MEDICAL SYSTEMS
D
IRECTION 2300749-100, REVISION 2LOGIQ
TM
3 SERVICE MANUAL
Chapter 3 3-47
3-11-4 Upgrading Application Software (cont’d)
If possible, while the system is rebooting remove the CD from the drive. If you didn’t do that don’t worry
you will get a dialog box like the one shown in Figure 3-31.
9.) REMOVE the Applications CD
10.)From the Start Applications dialog box, select START
.
11.)Select CANCEL
.
3-11-5 Possible Behavior of the First Power Up After Application Load
After installing the new Application Software, if a new Front Panel code is needed:
The first time the Ultrasound application (Echoloader) is run after the software upgrade, the application
will detect that the Front Panel code is out of date. Because the Front Panel has to have power removed
and re-applied before the Front Panel code update process can continue, the application boot up will
automatically shut down the LOGIQ
™ 3 just before the application boot up is complete.
In some cases, the system may not shutdown automatically. If the dialog box appears, "Ending Process,
Please wait...", please wait 2 minutes. If the system does not turn off, force the shut down by holding
down the Standby switch until the monitor turns off then let it complete the power down process.
Simply ensure the system has powered down, and the power button is amber, then restart the system
to load the Ultrasound application. The auto-power down will only occur on the first power up after the
application upgrade.
Figure 3-31 Start Software