3-11-4 Upgrading Application Software
1.) Disconnect the system from the network.
2.) Place the “Application Software” CD-ROM into the CD-ROM drive.
3.) Power Down the scanner.
4.) Power up the scanner. Wait for a dialog box like the one in Figure 3-27.
5.) Select INSTALL SW ...
.
6.) There will be two dialog boxes popping up warning you that you are about to install new
software as shown in Figure 3-28. In both cases click OK
.
WARNINGWARNING
While the software install procedure is designed to preserve data, you should save any
patient data, images, system setups to a CD-R or MOD or hardcopy.
CAUTION
Remember to remove the CD-R or MOD before continuing with the software install.
Figure 3-27 Start Application Window
Figure 3-28 StartLoader Dialog Boxes