3-10-4 Base Load Software Load Procedure For BEP Ver 1.0
NOTE: Disconnect the system from the network and remove all transducers, all External Drives (USB
MOD, Flash Sticks, USB Memory etc.,), Switch Off peripherals like (VCR, VGP) before
installing Base Load.
1.) Insert the disk labeled “Base System Software Load Image as per BEP Type” into the CDROM
drive and switch on the system.On the screen that appears, Press any key to Continue.
2.) You Will See LOGIQ 3 Base Installation screen, It will ask for the 6 digit ID, Enter any 6 digits
(e.g, 111111).
3.) Enter ‘A’ has the Installation Type.
Figure 3-21 Base Installation Screen