Service manual

GE MEDICAL SYSTEMS PROPRIETARY TO GE
D
IRECTION , REVISION
3-36 Section 3-10 - Loading Base System Software
3-10-3 Base Load Software Load Procedure For BEP Ver 2.0 (cont’d)
6.) Remove the CD-ROM
NOTE: Remove the CD-ROM from the drive; otherwise you will be repeating the Base System
Software Load process.
7.) After removing the CDROM from the drive, properly turn off the scanner by pressing and
holding power on button on keyboard for more than 20 seconds.
8.) Turn the scanner back on. It will now boot up and automatically log on to start checking
hardware and Device Manager. This is a normal and should be allowed to run to completion.
NOTE: When booting up the first time it could take 30-90 seconds after the login box appears before
the trackball and key board are active. Watch for the arrow cursor to appear
.
NOTE: While the script is running, several windows or dialog boxes will appear on the screen. Wait for
the “System Settings Change” dialog box “Restart computer now” (approximately 3 minutes
after the desktop). Do NOT touch the system during this process. Activating the keyboard,
mouse or front panel could corrupt the installation.
NOTE: At times it may look like the system is unresponsive. PATIENCE! The process takes 3-4
minutes. Wait for the “Restart computer now” message to appear before activating the
keyboard, mouse or front panel controls.
9.) When you get a message to reboot Windows, wait 15 seconds or until the hour glass
disappears and select NO
.