User guide

Model 3i Digital Force/Torque Indicator User’s Guide
20
12 SPECIFICATIONS
12.1 General
Accuracy:
±0.2% of full scale ±1 digit + sensor
Sampling rate:
2,000 Hz
Power:
AC or rechargeable battery. Low battery indicator appears when battery level is low, and
indicator powers off automatically when power reaches critical stage.
Battery life:
Backlight on:
up to 7 hours of continuous use
Backlight off:
up to 24 hours of continuous use
Measurement units:
lbF, gF, kgF, N, kN, lbFin, ozFin, kgFm, kgFmm, Nm, Ncm, (depending on sensor)
USB output:
Configurable up to 115,200 baud
Configurable
settings:
Digital filters, USB output, set points, automatic shutoff, default units, default mode, key
tones, audio alarms, calibration
Safe overload:
150% of full scale (display shows “OVER” at 110% and above)
Weight:
0.7 lb [0.3 kg]
Included accessories:
Carrying case, AC adapter, battery, USB cable, resource CD (USB driver, MESUR
TM
Lite
software, MESUR
TM
gauge DEMO software, and user’s guide), NIST-traceable certificate
of calibration
Warranty:
3 years (see individual statement for further details)
12.2 Factory Settings
Parameter Setting
Set points
Upper Disabled (defaults to 80% of full scale, compression, when enabled)
Lower Disabled (defaults to 40% of full scale, compression, when enabled)
Filters
Current 8
Displayed 1024
Backlight Auto
Minutes 1
USB Output
Baud Rate 9,600
Data Format Numeric + units
Automatic Shutoff Enabled
Minutes 5
Beeps
Keys Enabled
Alerts Enabled
Set Points Momentary
LCD Contrast 10
Initial Mode Real Time
Units Depends on sensor