Items you can change:
Timeout Enter the timeout in milliseconds for the input to
timeout. If a new Data Provider packet does not arrive
within this time the Timeout Action will trigger.
Generally this timeout should be set to at least three
times the acquisition module transmission rate.
Timeout Action Select the action to take place when a timeout occurs.
i.e. when communications (for more than the duration
of the Timeout value) is lost with the acquisition
module.
See the Output Actions section for the available
actions and the effect of these choices on the different
output ranges.
Remote Error Action Acquisition modules can report errors.
See the Output Actions section for the available
actions and the effect of these choices on the different
output ranges.
Remote Batt Action When the acquisition module reports a low battery this
action will occur. See the Output Actions section for
the available actions and the effect of these choices on
the different output ranges.
Output Actions
The following actions can be selected.
None Do nothing
Minimum Full Scale Set analog output to the minimum full scale value
Maximum Full Scale Set analog output to the maximum full scale value
Minimum Output Set analog output to the minimum possible value
Maximum Output Set analog output to the maximum possible scale value
Half Full Scale Set analog output to halfway between minimum and
maximum full scale value
Hold Last Output Hold the last output. (Does the same as None for the
Timeout Action)