CF 183 16 32-1113 REV 2 1010
PC Peak Compression mode for primary reading
CLR Clear peaks
Z Zero display and perform the CLR function
Filters
FLTCn Digital filter for displayed readings
FLTPn Digital filter for current readings
n= 0-10, filter = 2
n, ex: n=0= no filter, n=10=1024 samples averaged
Memory & Statistics
MEM Transmit all stored readings
STA Transmit statistics
Set Points
SPHD Disable high set point
SPLD Disable low set point
SPHn High set point. n=value (+ for compression, - for tension)
SPLn Low set point. n=value (+ for compression, - for tension)
Note: High set point value must be greater than low set point value if both values are set to the same
polarity.
USB/RS-232 Communication
FULL USB/RS-232 transmission with units
NUM USB/RS-232 transmission without units (only numeric values)
AOUTn Auto-transmit n times per second n=1,2,5,10,25,50,125,250. 0=disabled
Note: n = 1 = yields 50 times per second. This is provided for backward compatibility with legacy
gauges.
Mitutoyo Communication
MIT Enable Mitutoyo output
MITD Disable Mitutoyo output
POL Mitutoyo output with polarity (+ for compression, - for tension)
NPOL Mitutoyo output without polarity (absolute value)
PM Print/send data to a Mitutoyo-compatible device
Averaging
A Enable Average mode
AD Disable Average mode
AM Select Average mode (if enabled) for primary reading
ATn Average time. n=0.1-300.0 seconds
DELn Initial delay. n=0.1-300.0 seconds
TRFn Trigger force. n=value (+ for compression, - for tension)
19
External Trigger
ETH Enable high level-triggered External trigger mode
ETL Enable low level-triggered External trigger mode
ETHL Enable reading captured on a high to low transition
ETLH Enable reading captured on a low to high transition
ETD Disable External trigger mode
Input / Output Bits
Sn Set output bit (open drain, pull to ground). n=0,1,2
Cn Clear output bit. n=0,1,2
Rn Read current status of output bit or level of input pin. n=0,1,2,3
Personality
RN Read product name
RM Read model number
RV Read firmware version number