Manual

CF10 33 6/24/08 Rev. H
6.1 CALIBRATION GUIDELINES
The following guidelines should be considered when calibrating a sensor:
Apply a calibration load that approximates the load to be applied during system use, using dead weights or a
testing device (such as an MTS or Instron). If you intend to use a "puck" during testing, also use it when
calibrating the sensor. See Sensor Loading Considerations for more information on using a puck.
Avoid loading the sensor to near saturation when calibrating. If the sensor saturates at a lower load than desired,
adjust the "Sensitivity".
Distribute the applied load evenly across the sensing area to ensure accurate force readings. Readings may vary
slightly if the load distribution changes over the sensing area.
Note: Read the Sensor Performance Characteristics section before performing a Calibration.
6.2 CALIBRATION PROCEDURE
IMPORTANT! Sensors must be properly conditioned prior to calibration and use.
1. Select File -> New to open a new Real-time window. If you wish to use a previously saved calibration file,
select Tools -> Load Calibration, enter the calibration file name and path, and click OK.
2. Load each sensor with a known force.
3. Click Tools -> Change Sensitivity. If you have the Multi-Handle ELF system, adjust sensitivity for each
sensor, if necessary, by either dragging the slider up and down, or entering an integer between 1 and 20
(default is 12). The "Sensitivity" should be adjusted whenever necessary to keep the force display in a usable
range.