User Manual
REVB10/18/12,ECN8972
8
Currentplotdatacanbeexportedtoaselectedtab‐delimitedfileforconvenientimportingin
aspreadsheetorotherapplicationsuchastheVibraScoutPostProcessor.Thisfeature
providesanalternativetothedatalogfilebyallowingtheusertoexportdataafter
acquisition.
3. DeviceInfo
Model # and Serial # are stored in the device EEPROM and read when the VibraScout™
applicationislaunched.Devicememoryisalsoavailabletostoreuserinformation.Enterup
to180charactersintheUserInfofieldandclickonSaveUserInfotostoretheinformationin
non‐volatilememory.WhenVibraScout™islaunched,theinformationwillbereadfromthe
deviceandappearintheUserInfofield.
4. ApproximateBandwidth
Select the desired bandwidth based on the application requirements.The noise level
increaseswithbandwidth, soselectingalowerbandwidthwillprovideimprovedsignal/noise
at the expense of frequency response.The 7543A sample rate is equal to approximately
twice the selected bandwidth.In cases where data overflows occur due to inadequate
systemresources(CPUspeed,RAM),selectingalowerbandwidthwillusuallyeliminatethe
overflows.SeethesectionontheDataIntegrityindicatorformoreinformation.
5. DataFileSettings
TheDataFilepathisselectedwiththeBrowsebutton.Dataaresavedastab‐delimitedtext
for easy importing into a spreadsheet or other application when the Log To File box is
checked in the Acquisition screen. Software automatically appends date and time code to
theendofthefilenameforeasieridentificationoftestsequence.Datafileisclosedandthis
fieldisclearedeverytimeacquisitionisstopped.SelectingtheLog ToFileboxwillprompt
userforcreatinganewdatafile.
6. Offsets
The 7543A is calibrated at the Dytran factory along each orthogonal axis using precision
fixturesandearth’sgravity.Whenthe 7543A isplacedatanyrandomposition,theOffsets
Acquirebuttonallowsthevaluesforthethree accelerationaxespitchandrolltobecaptured
and set to zero.Therefore,
relative measurements may now be made from that position.