11.2 Mode
Mode is one byte of data (2 characters). It uses the prefixes G, R, and W with the suffix 07. This byte controls
the operating mode of the meter as follows:
0: Batch Mode
1: Rate Mode
2: Sq Rt Mode
The default value is 0, Batch Mode
EXAMPLE: To set the mode to Rate for meter #15, send the command *15W0701<CR>. The meter will
respond with 15W07<CR>. To reset the meter, send the command 15Z05<CR>. The meter will respond with
15Z05<CR>.
11.3 Display Mode
This item is 1 byte (2 characters). It uses command prefixes G and P with the suffix 33, allowing you to read or
change the meter’s display mode.
a) The bit pattern in response to a G33<CR> command is as follows:
Table 11.2. Display Mode Bits
FUNCTION BIT
NO.
VALUE
RATE AND SQ RT MODES BATCH MODE
0-2 0 Not used (Defaults to 1) Not used (Defaults to 1)
1 RATE is displayed BATCH is displayed
2 AVG RT is displayed No. of BATCHes is displayed
3 TOTAL is displayed TOTAL is displayed
4 TIME is displayed TIME is displayed
3-6 0 Not used Not used
1 Not used Not used
7 0 Not used RUN mode
1 Not used Stop mode
(meter is in STOP mode)
b) Bit pattern for a P33 command data format:
FUNCTION BIT
NO. VALUE
RATE AND SQ RT MODES BATCH MODE
0-2 0 Not used (Defaults to 1) Not used (Defaults to 1)
1 Puts RATE on display Puts BATCH on display
2 Puts AVG RT on display Puts No. of BATCHes on display
3 Puts TOTAL on display Puts TOTAL on display
4 Puts TIME on display Puts TIME on display
5, 6, 7 Not used Not used
3-7 Not used-Set to zero
Not used
Not used-Set to zero
Not used
EXAMPLE: In RATE MODE, you want to change the display, which is currently showing a rate value, to display
TOTAL on meter #15.
Send command: “15P3303<CR>”
Meter will send back this message: “15P33<CR>” and change its display to show “TOTAL”