Owner manual

DFI INFINITY CS (V- M3598/N/0906) 29 CF 70
ACCURACY AT 25
o
C
Max Error Strain: ±0.03% of reading, ± 1 count
Span Tempco: 50 ppm/°C
Step Response: 1 sec
Warm Up to Rated Accuracy: 30 min
Excitation Voltage: AC power units
DC power units
24 V @ 25 mA, Refer to Table 8-1
12 V @ 50 mA, Color chart for DC Output
10 V @ 120 mA, Excitation
5 V @ 60 mA
Load Regulation: 1.1%
Line Regulation: 0.02% per Vac
ALARM OUTPUTS (if applicable)
2 Form “C” on/off relays. Configurable for latched and unlatched by software.
Max current: 5 AMPS, resistive load
Max voltage: 250 Vac or 30 Vdc
ANALOG OUTPUT (if applicable)
Signal Type: Current or voltage
Signal Level: Current: 10 V max compliance at 20 mA output
Voltage: 20 mA max for 0-10 V output
Function: May be assigned to a display range or proportional control output with setpoint #1
when used as a control output.
Linearity: 0.2%
Step Response Time: 2 – 3 seconds to 99% of the final value
ISOLATED ANALOG OUTPUT (TB5, if applicable)
Same as non-isolated analog output except isolated.
Signal Type: Current or voltage
Signal Level: Current: 10 V max compliance at 20 mA output
Voltage: 20 mA max for 0-10 V output
Function: May be assigned to a display range or proportional control output with Setpoint #1
when used as a control output.
Linearity: 0.2%
Step Response Time: 2 – 3 seconds to 99% of the final value
Isolation: 130 Vrms working voltage, 1000 V/60sec Dielectric test
Only one analog output is available on each unit and it must be factory installed.
INPUT POWER INFORMATION
AC units 115/230 V~(AC) ±10%, 50/60 Hz
9.5 W max, power consumption (Non-Isolated Analog Out)
11.0 W max, power consumption (Isolated Analog Out)
DC Units 10-32 Vdc or 26-56 Vdc, 8 W
Do not use a combination of dc power and internal excitation or Isolated Analog
Out, unless using dc power of 20-32 Vdc.
Refer to Table 8-1 below
External Fuse Required:
IEC 127-2/111
Power Fuse
115V 125 mA @ 250 (T)
230V 63 mA @ 260 (T)
UL 248-14 (Listed Fuse)
Power Fuse
115V 175 mA @ 250 V Slow-Blow
230V 80 mA @ 250 Slow-Blow