Owner manual

DFI INFINITY CS (V- M3598/N/0906) 19 CF 70
= Setpoint 1 assigned to net reading (Default)
= Setpoint 1 assigned to gross reading
2. Press the S/NT/GRS button to toggle between available choices.
3. Press the MENU button to store your selection(s). The meter momentarily shows
, followed by
(Setpoint 2 Configuration).
4.7 Using Setpoint 2 Configuration
Refer to Table 6-1 for a summary list of menu configuration.
Setpoint 2 Configuration
is not active unless your meter has dual relay output capabilities. The LED’s will
display whether the
is active or not. You may use Setpoint 2 Configuration for the following:
To set setpoint’s active band above or below your chosen value
To select whether the setpoint operation is latched or unlatched
To assign setpoint values to the net or gross reading
4.7.1 Setting Setpoints 2’s Active Band
1. Press the MENU button until the meter shows .
or
2. Press the X/TARE button. The meter shows one of the following:
= (Active above the setpoint) (Default)
= (Active below the setpoint)
3. Press the S/NT/GRS button to toggle between available choices.
4. Press the X/TARE button to select if Setpoint 2 is latched/unlatched or press the MENU button to store your
selection. The meter shows
.
4.7.2 Selecting if Setpoint 2 is Latched or Unlatched
1. Press the MENU button until displays, then press the X/TARE button twice.
or
Press the X/TARE button from
.
The meter shows one of the following:
= Setpoint 2 to be unlatched (Default)
= Setpoint 2 to be latched
2. Press the S/NT/GRS button to toggle between available choices.
3. Press the MENU button to store your selection(s). The meter momentarily shows
, followed by
(Setpoint 1 Deadband).
4.7.3 Assigning Setpoint 2 Values to Net or Gross Readings
1. Press the MENU button until displays, then press the X/TARE button twice.
or
Press the X/TARE button from
.
The meter will show one of the following:
= Setpoint 2 assigned to net reading (Default)
= Setpoint 2 assigned to gross reading
2. Press the S/NT/GRS button to toggle between available choices.
3. Press the MENU button to store your selection(s). The meter momentarily shows
, followed by
(Setpoint 1 Deadband).
4.8 Setting the Setpoint 1 Deadband
Refer to Table 6-1 for a summary list of menu configuration.
Setpoint 1 Deadband
is not active unless your meter has dual relay output capabilities. The LED’s will
display whether the
is active or not. The Setpoint 1 Default deadband is 0003. To change the deadband
(hysteresis) of Setpoint 1, follow these steps: