Owner manual

DFI INFINITY CS (V- M3598/N/0906) 17 CF 70
You may use Reading Configuration to configure your meter for the following:
To select ratiometric or non-ratiometric operation
To set the input resolution of your meter
To display the filtered/unfiltered signal input value
To select gross/net vs. peak reading
4.4.1 Selecting Ratiometric/Non-Ratiometric Operation
1. Press the MENU button until displays.
2. Press the X/TARE button. The meter flashes one of the following:
(Ratiometric reading) (Default – for strain meters)
(Non-ratiometric reading – typically for voltage & current transducers)
3. Press the S/NT/GRS button to view the last stored selection and to toggle between selections.
4. Press the X/TARE button to select input resolution or press the MENU button to store your selection and
shows
menu.
4.4.2 Setting Input Resolution
To set the input resolution of your meter, follow these steps:
1. Press the MENU button until
displays, then press the X/TARE button twice.
or
Press the X/TARE button from
.
One of the following displays (default is
):
= 10µV Unipolar inputs. 25 µV for bipolar inputs.
= 1µv for Unipolar inputs. 5 µV for bipolar inputs.
= 2µV for Unipolar inputs. 10 µV for bipolar inputs.
= 3µV for Unipolar inputs. 15 µV for bipolar inputs.
= 5µV for Unipolar inputs. 25µV for bipolar inputs.
Example: 3 µV resolution means that if you input 0-30 mV, at 30 mV the meter shows
.
2. Press the S/NT/GRS button to scroll through available selections.
3. Press the X/TARE button to display the filtered/unfiltered signal input or press the MENU button to store your
selection and shows
menu.
4.4.3 Displaying the Filtered/Unfiltered Input Signal
To display the filtered/unfiltered signal input, follow these steps:
1. Press the MENU button until
displays, then press the X/TARE button three times.
or
Press the X/TARE button from
.
One of the following displays:
= (Filtered value) (Default)
= Unfiltered value
2. Press the S/NT/GRS button to toggle between available choices.
3. Press the MENU button to store your selections.
momentarily displays, followed by Menu.
4.4.4 Selecting Gross/Net or Peak Display
To select gross/net or peak display.
1. Press the MENU button until
displays, then press the X/TARE button three times.
or
Press the X/TARE button from
.
One of the following displays:
= Gross/Net Display (Default)
= Peak Display
2. Press the S/NT/GRS button to toggle between available choices.
3. Press the MENU button to store your selections.
momentarily displays, followed by Menu.