Manual

20.2 Ot.SC.OF Enter OUTPUT SCALE & OFFSET CONFIGURATION Submenu:
READING 1
Press ‘RESET/ENTER 10) Display shows "REAd 1" Reading 1 menu.
Press ‘RESET/ENTER 11) Display flashes 1st digit of previous "REAd 1" value.
Press ‘/MAX’ & ‘/MIN 12) Enter "REAd 1" value. (Example 000000).
Press ‘RESET/ENTER 13) Display advances to "OUt 1" Output 1 menu.
OUTPUT 1
Press ‘RESET/ENTER 14) Display flashes 1st digit of previous "OUt 1" value.
Press ‘/MAX’ & ‘/MIN 15) Enter "OUt 1" value. (Example 0000.00).
Press ‘RESET/ENTER 16) Display advances to "REAd 2" Reading 2 menu.
READING 2
Press ‘RESET/ENTER 17) Display flashes 1st digit of previous "REAd 2" value.
Press ‘/MAX’ & ‘/MIN 18) Enter "REAd 2” value. (Example 999999).
Press ‘RESET/ENTER 19) Display advances to "OUt 2" Output 2 menu.
OUTPUT 2
Press ‘RESET/ENTER 20) Display flashes 1st digit of previous "OUt 2" value.
Press ‘/MAX’ & ‘/MIN 21) Enter "OUt 2" value. (Example 0010.00).
Press ‘RESET/ENTER 22) Display Display shows "StOREd" stored message momentarily and then
advances to the "SP CNF" SetPoint Configuration menu.
The above example is for 0-10 V of the entire range of the process input and analog output. For 0-20mA output
you need to set "MOdE" to "CURRNt" DC-Current output and "OUt 2" to 0020.00.
21.0 SETPOINT CONFIGURATION MENU FLOWCHART