Manual

6.5.2 For ±50mV, 0-100mV ranges with External Excitation:
S1_L; S2_A/B/H, jumpers are closed as shown:
6.5.3 For ± 500mV, 1V ranges with Internal Excitation:
S1_L/E/F; S2_G/H, jumpers are closed as shown:
6.5.4 For ± 500mV, 1V ranges with External Excitation:
S1_L/E/F; S2_H, jumpers are closed as shown:
6.5.5 For ± 5V, 0-10V ranges with Internal Excitation:
S1_L/C/D; S2_G/H, jumpers are closed as shown:
6.5.6 For ± 5V, 0-10V ranges with External Excitation:
S1_L/C/D; S2_H, jumpers are closed as shown:
6.5.7 For ± 50V, 0-100V ranges with Internal Excitation:
S1_L/A/B; S2_G/H, jumpers are closed as shown:
6.5.8 For ± 50V, 0-100V ranges with External Excitation:
S1_L/A/B; S2_H, jumpers are closed as shown:
6.5.9 For 0-20mA or 4-20mA range with Internal Excitation:
S1_K; S2_A/B/C/D/G/H, jumpers are closed as shown with Internal Excitation: